Beam Analyzer Measurement Device
About the
Beam Analyzer Measurement Device
The Beam Analyzer is an award-winning device that allows you to measure the beam specifications of laser light show projectors up to 100W with incredible accuracy. The Beam Analyzer is essentially a “lie detector” that helps you determine if your laser’s beam specifications are correct or not.
The Beam Analyzer allows you to measure in real time:
- Beam Divergence
- Beam Alignment (X*Y mrad)
- Power Output
- Lens collimating of laser sources
Optical specifications
Symbol | Min | Typ | Max | Unit |
---|---|---|---|---|
Optical power [Popt] | 0.1 | - | 100 | W |
Power measurement accuracy | - | 1 | - | % |
Resolution on divergence [RΦ] | - | 0.01 | - | mrad |
Divergence measured (FWHM) [Φ] | 0.1 | - | 10 | mrad |
Input beam diameter [D] | - | - | 22 | mm |
Wavelength detected [λ] | 400 | - | 1000 | nm |
Mechanical specifications
Parameter | Symbol | Value | Unit |
---|---|---|---|
Length x Width x Height (without the support / handle) | L x W x H | 100 x 75 x 57 | mm |
Weight | msys | 500 | g |
Electrical specifications
Parameter | Symbol | Value | Unit |
---|---|---|---|
Typical consumption (USB 3.0) | PUSB | 2 | W |
(*) The sensor of the LBA-C must remain below 75°C while operating. It may be need to dissipate heat from the case with a fan when using the device with high power beams for a long period.
(**) A correct alignment should be done before setting the laser projector to maximum power. The user must follow the alignment procedure described in the User Manual to prevent damage to the system.